6.0 CALIBRATION

 This parameter allows the user to calibrate the scale any
time during the operation.
 Switch the scale off and then on it again by pressing
[On/Off]. During the self-test press [Smpl] and [Print]
together.
 Display will show "unLoAd". Remove all weights from the
pan.
 Wait for the stability indicator to be on and then press [Zero].
 The display will then show the first calibration weight request.
 Press [Mode] to scroll through the other calibration weights.
 Press [Zero] to select calibration weight. The display will
show "LoAd".
 Place this weight on the platform and press [Zero] when the
stability indicator will be on to show the value is stable.
 If the calibration is complete, the scale will show "PASS" and
will return to normal weighing counting back to zero. Remove
the calibration weight at this stage.
 If the wrong calibration weight is used, the scale will show
"FAlL H" (when the weight is higher) or "FAlL L" (when the
weight is lower). Repeat the process using the correct
calibration weight.
Calibration weights:
Model #
ACBplus
150
Weight 1
50g
Weight 2
100g
Weight 3
150g

7.0 COMMUNICATION WITH A COMPUTER/PRINTER

The ACBplus series of scales come with a standard RS-232 interface.
The standard Interface parameters are:
RS-232 output of weighing data
ASCII code
4800 Baud
8 data bits
No Parity
© Adam Equipment Company 2004
ACBplus
ACBplus
300
600
100g
200g
200g
400g
300g
600g
ACBplus
ACBplus
600H
1000
200g
500g
400g
1000g
600g
-
Connection details are:
Connector: 9 pin d-subminiature socket
Pin 3
Output
Pin 2
Input
Pin 5
Signal Ground
17
ACBplus
ACBplus
1500
3000
500g
1000g
1000g
2000g
1500g
3000g
ACBplus
5000
2000 g
5000 g
-