Information About Errors - Nissei DSK-1031 Instructoin Manual

Hide thumbs Also See for DSK-1031:
Table of Contents

Advertisement

Saved values are kept with the measurement date and time, if set. If you want to save
time and date together with the measurement values, date and time should be set
before the measurement.
When an error occurs (ERR), the measurement results are not saved.
Viewing the Saved Data
1. Turn off the device by pressing the START/STOP
button. To view the results stored in the memory
unit 1, press , and to view the results stored in the
memory unit 2, press . The selected memory unit
will be underlined on the display.
The display shows the average value of stored
results indicated with index « » (Fig. 19). Average
value is displayed when the selected memory unit
contains two or more saved measurement results.
2. Each time you press the
button or
saved measurement results will be displayed in a
sequential manner.
3. Memory unit number, measurement date and
time will be displayed one by one in the upper right corner of the display.
4. The result saved in the memory unit 1 is the most recent among the saved data in
the selected memory. The higher memory number, the older the result.
The memory data are displayed for about 30 seconds and after that, if no button is
pressed, the device is automatically turned off .
Pressing the
button switches the display from records in the memory unit 1 to
records in the memory unit 2, while pressing the
recorded in the memory unit 1.
5. Press the START/STOP button to turn off the device.
Deletion of Saved Data
Readings can be deleted unit by unit: all together or separately. The memory may be
cleared when the selected memory unit contains two or more results.
1. Select a value from the memory unit to be deleted or the average value (unit with
index « ») to clear the whole memory unit.
2. Press and hold the
button or
Selected memory
indicator
button,
Number of saved measurements
Fig.19
button results in returning to data
until the sign «---» appears on the display.

INFORMATION ABOUT ERRORS

INDICATION
Blood pressure is too
The cuff is not at the heart level.
low or too high.
The cuff is put on incorrectly.
During the measurement, a person
Average
was talking or moving.
value
index
Measurement results
Eff ect of measurement conditions,
are diff erent each time.
physical or mental state.
Measurement results
Eff ect of relaxed state at home and
are diff erent in clinic
tension in clinic.
and at home.
Infl ation is repeated.
Pressure can not be measured due to
movement or talking.
Maximum allowable pressure:
pressure can not be measured due
to movement or talking during
the measurement, in spite of the
fact that the cuff was pumped at
maximum.
Pressure can not be measured due to
movement or talking.
or
Cuff is not securely connected to
the device.
Cuff is not put on properly.
No time indication on
Time was not set.
the display.
Note: there is no time indication when
batteries and/or power source are not
provided.
8
LIKELY CAUSE
METHODS OF CORRECTION
Put on the cuff at the heart level.
Check the cuff position on the arm.
Be calm and quiet during the
measurement.
Take measurements under the same
conditions.
Show the pressure records made at
home to your doctor for advice.
Re-infl ation is not a malfunction of
the device. Do not talk and do not
move during the measurement.
Do not talk and do not move during
the measurement.
Do not talk and do not move during
the measurement.
Check the connection.
Make sure that the cuff is put on
correctly.
Set time and date.
Install batteries and connect the
power source.