Date and time are
Time was not set or measurements
displayed as «--/--».
were carried out before setting the
time.
Batteries are discharged.
Nothing is shown on
Batteries are discharged.
the display.
Batteries are installed incorrectly.
Connecting terminals are
contaminated.
Power source is not connected.
Insuffi cient contact when touching
the button.
You touched the START/STOP button
when installing the batteries.
If, despite the above-given recommendations, you fail to obtain the right measurements, stop
the operation and contact the service center (addresses and telephone numbers of authorized
organizations are provided in the warranty certifi cate). Do not attempt to adjust the device
internal mechanism on your own.

WARRANTY

1. The manufacturer guarantees the warranty period of 5 years for the device from
the sale date provided that the consumer observes operation, transportation and
storage requirements. The warranty period for the cuff and the power source is 12
months from the sale date.
2. Warranty liabilities are documented with the warranty certifi cate upon selling the
device to the buyer. The guarantee is valid provided that the device has not been
opened or damaged by the buyer.
3. Addresses of organizations engaged in the warranty service are specifi ed in the
warranty certifi cate.
Set date and time.
Date and time can not be saved
without setting.
Replace all batteries with new ones.
Replace all batteries with new ones.
Install batteries properly.
Wipe connecting terminals with a
dry cloth.
Connect the power source.
Ensure stronger pressing the button.
Turn off the device by pressing the
START/STOP button and perform the
measurement again.
9

TECHNICAL SPECIFICATIONS

Operating Principle
Oscillometric method
Indicator
15 digits liquid crystal display
Pressure Indicating Range , mmHg 0-300
Measuring Range :
cuff pressure, mmHg
50-250(systolic), 40-180(diastolic)
pulse rate, bpm
40-160
Accuracy:
±3
cuff pressure, mmHg
±5
pulse rate, %
Infl ation
Automatic infl ation (air pump, Fuzzy Infl ation algorithm)
Defl ation
Automatic Defl ation (electric valve)
Electric supply voltage, V
6
Electric supply type
Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor
Memory
2 units, 60 values per each + average value
ADP-W5 Power Source
Output voltage, V
6
Maximum load current, A
0,5
Input voltage, V/Hz
100-240/50
Operating Condition
temperature, ˚C
+10 to +40
humidity, % Rh
15 - 85
Transportation / Storage Condition
temperature, ˚C
-20 to +60
10 - 95
humidity, % Rh
Cuff Size
adult (circumference of the upper arm 22-42 cm)
Overall dimensions:
Size (electronic unit), mm
115 x 115 х 65
Weight (without packaging,
bags, batteries and power
source), g.
406
Manufacture year:
is specifi ed on the device housing (in the battery
compartment) in the device serial number after the
letters «SN»