BLOOD PRESSURE MONITOR
INSTRUCTIONS
This manual is intended to assist you in the safe and efficient operation of BLOOD PRESSURE MONITOR DSK-1011J. The product must be used in accordance with the procedures contained in this manual and must
not be used for purposes other than those described herein. It is important to read and understand the entire manual. In particular, please read carefully and become familiar with the section entitled "TIPS ON TAKING
YOUR BLOOD PRESSURE".
C O N T E N T S
PA RT NAM ES A N D P R O D U CT COM P ON E NT S . . . . . . . . . . . . . 1
G E N E R A L I N F O R M AT I O N . . . . . . . . . . . . . . . . . . . . . . . . . 1
P R E C A U T I O N S A N D C O N T R A I N D I C AT I O N S . . . . . . . . . . . . 2
T I P S O N TA K I N G Y O U R B L O O D P R E S S U R E . . . . . . . . . . . . . 2
P R E PA R AT I O N B E F O R E U S E . . . . . . . . . . . . . . . . . . . . . 2
P A R T N A M E S A N D P R O D U C T C O M P O N E N T S
AC CONNECTOR
MEMORY KEY [1]
"SET" KEY
BATTERY
COMPARTMENT
AIR CONNECTOR
MAIN UNIT
G E N E R A L I N F O R M A T I O N
INDICATIONS FOR USE
This product is intended for noninvasive measurement of systolic and diastolic blood pressure,
determination of pulse rate and calculation of pulse pressure in adults in a home healthcare environment.
The product is not designed for neonatal use. Please consult with your doctor or physician to use this
product to take blood pressure of child or person in pregnancy or under pre-eclamptic condition.
WHO CLASSIFICATION SYMBOL
Measured blood pressure values
DISPLAY
are
classified
against
WHO
guideline (2009).
The higher class applies when
systolic and diastolic pressures fall
into different classes. For specific
information on your blood pressure,
contact your physician. Never make
any judgment on your own based
on the classification indication only.
WHO: World Health Organization
PULSE PRESSURE DISPLAY
This product calculates and displays pulse pressure.
While systolic blood pressure keeps increasing with age, diastolic blood pressure tends to start
decreasing at around age 50. Pulse pressure is value subtracting diastolic value from systolic value and
therefore pulse pressure tends to increase with age. Pulse pressure has been considered to be related to
stiffness of artery and has been studied as one of circulatory risk factors. It is sometimes said that 45
mmHg of pulse pressure is normal value. Pulse pressure only does not represent arterial sclerosis.
However, observation of pulse pressure in long term will be important.
CARE AND MAINTENANCE
Keep the product clean. Inspect its cleanliness after use. To clean, use only a soft dry cloth. Do not use
gasoline, paint thinner, or other strong solvents. Since the cuff may absorb perspiration and other fluids, inspect
it for stain and discoloration after each use. When cleaning the cuff, use a synthetic detergent and gently rub
the surface. Air dry thoroughly. Make sure fluid never gets in AIR HOSE. Do not machine wash or scrub it.
When storing the product, do not place heavy objects on it. Do not fold or bend the cuff forcibly. To
disconnect the cuff from the main unit, do not pull AIR HOSE but hold AIR PLUG and unplug it. Do not
coil AIR HOSE too tightly. When the product has been stored at a temperature below the freezing point,
keep it for at least 1 hour in a warm place before using. Remove the batteries if the product is to be
stored for an extended period of time. Keep the batteries out of reach of children.
We suggest that you have your blood pressure monitor checked every 2 years. This operation may only
be performed by the manufacturer or by firms authorized by the manufacturer.
METHOD OF MEASUREMENT
This product employs the oscillometric method for measurement of blood pressure and pulse rate. The
cuff is connected to the main unit and wrapped around the arm. Circuits within the cuff sense the small
oscillations in pressure against the cuff produced by the expansion and contraction of the arteries in
the arm in response to each heart beat. The amplitude of each pressure waves is measured,
converted to millimeters of mercury, and displayed on the LCD as a digital value.
MEASUREMENT ON INFLATION
Initially the cuff is rapidly inflated to the prescribed value at the
DSK-1011J
MEMORY KEY [2]
"START/STOP" KEY
DISPLAY
WHO CLASSIFICATION
SYS (mmHg)
DIA (mmHg)
Grade 3 Hypertension
>180
>110
Grade 2 Hypertension
160-179
100-109
Grade 1 Hypertension
140-159
90-99
High Normal
130-139
85-89
Normal
120-129
80-84
Optimal
<120
<80
M A K I N G A M E A S U R E M E N T . . . . . . . . . . . . . . . . . . . . . . . . 3
V I E W I N G T H E S AV E D R E A D I N G S ( M E M O R Y F U N C T I O N ) . . . 4
T R O U B L E S H O O T I N G . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
W A R R A N T Y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
AIR PLUG
AA (LR6) BATTERIES
AIR HOSE
CUFF
ATTACHED DOCUMENT
beginning of a measurement, during which the inflation mark is
displayed. After the pressure reaches this point, the inflation mark
disappears and the cuff is inflated more gradually, during which
pulse wave is detected. As pulse wave is detected, heart mark
flashes with synchronizing beep. Measurement ends when systolic
blood pressure is determined and air is exhausted from the cuff.
PERSONAL MODE
The inflation is adjusted based on the latest three readings in the
selected bank, either [1] or [2], to shorten the measurement time
and to improve comfort.
For example, high diastolic blood pressure values from past three
measurements will lead to inflation to higher pressure value than
the default value. In the event that this value is too high for your
current blood pressure, the cuff will rapidly deflate and
measurement will restart.
To use the personal mode effectively, chose the memory bank before starting a measurement. This
monitor remembers the last memory bank that was displayed. To choose the memory bank before
starting a measurement, display a reading saved in the desired memory bank, turn off the monitor
once and then start measurement.
BODY MOTION SYMBOL
Blood pressure value taken while moving cannot be said to be the correct value because body
movement can affect blood pressure.
This product analyzes pulse wave and displays [
[
] indicates the results might be affected by body movement.
IRREGULAR PULSE RHYTHM SYMBOL
Pulse rhythm can be disturbed from talking, moving or
arrhythmias. This product displays [
difference of shortest and longest time intervals between
pulse waves is over 25%, indicating irregular pulse
rhythm.
MEMORY FUNCTION
The measured values are automatically saved for later review in either one of two memory banks, [1]
or [2]. These two memory banks can be used to save readings of two persons separately or to save
morning readings and evening readings separately.
Each bank can save up to 60 readings. When the number of stored readings reaches 60, the oldest
reading will be deleted to record new reading.
Saved readings are recalled with measurement date and time when clock function is activated. *The clock
needs to be activated before measurement for date and time to be saved with the measured valued.
[Err] results are not saved.
PRODUCT SPECIFICATIONS
Model
: DSK-1011J
Operating Principle
: Oscillometric method
Indicator
: 15 digits liquid crystal display
Pressure Indicating Range
: 3 to 300 mmHg (cuff pressure)
Measuring Range
: 50 to 250 mmHg (systolic), 40 to 180 mmHg (diastolic), 40 to 160 bpm (pulse rate)
Accuracy*
: ±3 mmHg (cuff pressure), ±5% of reading (pulse rate)
Inflation
: Automatic inflation
Exhaust
: Automatic exhaust valve
Power Supply
: Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor ADP-W5 series
-1-
For test use
JAPANESE
ENGLISH
INSTRUCTIONS
PRESSURE
GRADUAL
INFLATION RAPID
RAPID
INFLATION
PRESSURE
AN
EXAMPLE
CHANGE
DETERMINED
"PERSONAL MODE"
] when body motion is detected.
regular pulse rhythm
] when the
An example of irregular pulse rhythm
compared to regular pulse rhythm
CHINESE
TIME
EXHAUST
INITIAL PRESSURE
CHANGE
TIME
OF
PRESSURE
WITH

Advertisement

loading

Summary of Contents for Nissei DSK-1011J

  • Page 1 INSTRUCTIONS This manual is intended to assist you in the safe and efficient operation of BLOOD PRESSURE MONITOR DSK-1011J. The product must be used in accordance with the procedures contained in this manual and must not be used for purposes other than those described herein. It is important to read and understand the entire manual. In particular, please read carefully and become familiar with the section entitled “TIPS ON TAKING YOUR BLOOD PRESSURE”.
  • Page 2 Protection against electric shock : Internally powered equipment, Type BF applied part (cuff) Power Consumption : 4W (max.) Mode of operation : Continuous operation Electric Rating : DC6V/4W, with four LR6 batteries Classification : Internally powered equipment Adaptor; AC100-240V, 50-60Hz, 0.12A, Key to symbols : Type BF applied part Unit;...
  • Page 3 Install four “AA” type batteries into the compartment. Touch “SET” KEY and keep your Make sure that the polarities correspond to the (+) and (–) finger on the key until “20 11” marks inside the battery compartment. starts flashing. AA (LR6) Batteries can be easily installed or removed as their (-) ends Keep your finger are pushed against the spring.
  • Page 4 This warranty does not cover damage or malfunctions caused by improper handling or use contrary to the instructions in this manual. Please contact your dealer for additional warranty provisions which may remain effective after the manufacturer's warranty period has expired. Manufacturer : NIHON SEIMITSU SOKKI CO., LTD. 2508-13 Nakago Shibukawa Gunma 377-0293 Japan web site http://www.nissei-kk.co.jp/english/...