29 Flushing and cleaning programs

29.1 When to use cleaning programs

Recommended cleaning intervals
Icon
User's manual
We recommend that you run the flushing and cleaning programs as follows.
Cleaning program
Short flushing
Long flushing
Suction flushing
Suction cleaning
Waterline cleaning
When
After every patient
In the morning and after the
working day
In the morning
After the working day
Continuous
When
cleaning
needed. We
enabled
recommend
that waterline
cleaning be
performed
once a month
after the
working day
(but not over
the weekend).
Continuous
When
cleaning
needed. We
disabled
recommend
that waterline
cleaning be
performed
once a week
after the
working day
(but not over
the weekend).
29 Flushing and cleaning programs
Duration
30 seconds / instrument
2 minutes (default; can be
programmed)
< 3 minutes
< 5 minutes
Cleans the dental unit's
waterlines.
The amount of disinfectant
concentrate used in a
cleaning cycle depends on
whether continuous cleaning
is enabled or disabled. For
more information, see table
Water and waterline
disinfection in section
"Technical specifications" on
page 273.
Duration: < 10 minutes in the
evening + min. 8 hours affect
time + < 30 minutes in the
morning
Planmeca Sovereign Classic