Download Print this page
Hide thumbs Also See for Sovereign Classic:

Advertisement

Quick Links

Planmeca Sovereign
Maintenance and infection control
After every patient
Surfaces
Wipe with Planmeca approved surface
1
disinfectant:
• metal parts of headrest
• metal parts of armrests
• protective fascia of the operating light
• handles of the operating light
• hygienic membrane
• instrument hoses
• instruments (see instrument instructions)
• instrument console*
• balanced instrument arms
• tray table
• cuspidor top cover
Wipe the patient mirror of the operating light
with a mild soap- and water solution.
Wipe the upholstery with a mild soap- and
2
water solution. Wipe with Planmeca approved
upholstery disinfectant if needed.
Wipe the monitor and the handle (see monitor
3
instructions).
Clean the bowl:
4
• Pour a few drops of Planmeca approved
suction disinfectant into the bowl and use
a soft brush.
• Rinse the bowl by pressing the bowl
rinse button.
• Wipe the outside of the bowl.
Suction system
Remove used aspirating tips.
1
Rinse the suction tubes by aspirating one
2
glass of water with each suction tube.
Wipe the suction handpieces with Planmeca
3
approved surface disinfectant.
Wipe the Flexy™ holder with Planmeca
4
approved surface disinfectant.
Waterlines
Short flushing
Place all water consuming instruments
1
into the flushing holder.
Bend the instrument arms to an angle of
2
at least 90° and start short flushing.
Classic
®
1
2
*Before cleaning the instrument console, lock the touch screen.
Touch screen is unlocked. Press to lock.
Touch screen is locked. Press and hold to unlock.
2
➁ 1
Do not
Do not
spray!
spray!
3
3
4
➁ 2
4
Do not
spray!

Advertisement

loading

Summary of Contents for Planmeca Sovereign Classic

  • Page 1 Wipe the monitor and the handle (see monitor instructions). Clean the bowl: • Pour a few drops of Planmeca approved suction disinfectant into the bowl and use a soft brush. *Before cleaning the instrument console, lock the touch screen. • Rinse the bowl by pressing the bowl rinse button.
  • Page 2 In the morning Suction system Suction flushing Start suction flushing. Wipe the suction handpieces with Planmeca ➁ 2 approved surface disinfectant. Filters -M IN Clean the coarse filters. -MA X Waterlines -M IN Long flushing Place all water consuming instruments into the flushing holder.
  • Page 3 Weekly Surfaces Wipe the upholstery with Planmeca approved upholstery treating agent. Suction system Mix 20 ml of Planmeca approved suction disinfectant with 1 l of water. Press the VS/A button. The bowl valve opens. Note, Dürr amalgam separators or VS/A systems only. Pour the mixture into the bowl.
  • Page 4 3 party device manufacturers. For detailed instructions, please see the user’s manual. Copyright Planmeca. Rights for changes reserved. www.planmeca.com...