29 CLEANING / PREVENTIVE MAINTENANCE

29 CLEANING / PREVENTIVE MAINTENANCE

29.1 Scanner tip

29.1.1 Cleaning
29.1.2 Chemical disinfection
146
Planmeca Chairside CAD/CAM solution
To guarantee the scanner's proper operation, it must be
checked and serviced once a year by a qualified
Planmeca service technician.
NOTE
For intraoral scanning systems only.
The following instructions are for the removable tip of the
scanner, not for the entire scanner. See instructions
below for cleaning the base of the scanner.
WARNING
Do NOT autoclave tips with no the temperature sym-
bol.
WARNING
Do NOT place in ultrasonic cleaner.
Use a disinfectant wipe to remove any visible debris on
the scanner tip
and/or
rinse the tip for 2 minutes under running tap water to
remove the rest of the debris.
After cleaning continue with chemical disinfection
autoclaving, see the sections 29.1.2 "Chemical
disinfection" and 29.1.3 "Autoclaving"below.
Before performing chemical disinfection clean the tip as
instructed in section 29.1.1 "Cleaning" above.
1. Immerse the tip in chemical disinfectant for manufacture
recommended disinfection time.
2. Remove the tip from the chemical disinfectant.
3. Rinse the tip according to the disinfectant manufacturer's
instructions.
4. Wipe water off the mirror using optical wipes.
5. Pack individually (for example in an autoclave pouch) and
store for later use.