4.2.6 DATALOGGER operating mode.

The LOGGER function permits to automatically measure the level, the
Carrier/Noise ratio and the Video/Audio ratio (the latter only in the case of analogue
channels) as well as the power channel measurements of each active channel in the
channel plan (see paragraph Channel Plan editor at section 4.2.1. Unit Global
Configuration). These measurements are stored in the memory so that they can be
subsequently viewed, printed or transferred to a PC. The
up to 55 obtained channels or loggers in the memory, with a maximum of up to 140
analysed channels in each.
When pressing the LOGGER
[16] key, a screen similar to the one of
the attached figure will appear. The first
line shows the logger number
attached figure) followed by the date it was
acquired (only if MEASURE function was executed
on that logger previously). Below, the different
functions you can perform from this operating mode
are shown: VIEW, MEASURE a
On the bottom of the display appears present time and date. To modify them
see section 4.2.1 Global Configuration Menu.
To accede to any of the different fields of the initial screen of the LOGGER
function, turn the rotary selector [9] until it is activated (it appears shadowed) and next
press it.
First of all you must select the logger on which you want to perform any
function: turn the rotary selector until the LOGGER NUMBER field is shadowed and
press it. Then turning the rotary selector, select the desired logger (from 0 to 54) and
finally press it again to validate. The acquisition date appears under the logger number
(only if you have taken measurements on this logger).
To carry out a logger acquisition you must select the MEASURE function, for this
purpose turn the rotary selector [9] and when this field appears shadowed press it, in this
way you will accede to the logger. Next press again the rotary selector [9] in order that the
PROMAX-8
SE
makes all the measurements defined in the LOGGER configuration
menu over all the active channels of the channel plan (see Edit Channel Plan function at
section 4.2.1 Global configuration menu).
The processing of channels as analogue or digital and the parameters to make
the measures (the frequency of the audio carrier for the analogue channels and
the frequency offset for the noise measure of the digital channels C/N ratio) will
correspond with the configuration of the equipment at the time of carrying out the
measurement.
To return to the initial menu of the LOGGER function press the LOGGER key.
09/2013
SE
(i.e. 53 in the
nd PRINT.
PROMAX-8
SE
will store
Figure 43.- LOGGER function.
Initial menu.
IMPORTANT
Page 37