Error Codes - NSK Primado2 Operation Manual

Total surgical system
Hide thumbs Also See for Primado2:
Table of Contents

Advertisement

8-2-3 Washing and disinfecting
The Motor can be washed via Thermo Disinfector.
If a Thermo-Disinfector is not available, proceed to 8-2-4 Sterilization below.
The following conditions are recommended for Thermo Disinfector:
(It is recommended to use a thermo disinfector compatible with ISO 15883)
a. Cleaning: at 45°C for 3 minutes
b. Intermediate rinse: At least 1 minute (using distilled, deionized or demineralized water)
c. Thermal disinfecting: at 93°C for 5 minutes (using distilled, deionized or demineralized water)
d. Drying: at least 10 minutes
Prior to use of a Thermo Disinfector, read the Thermo Disinfector Operation Manual carefully.
CAUTION
After cleaning and prior to sterilization, check the Motor in the same method as in Step "4. Check before Operation".
Do not use a chlorinated cleaner.
Do not clean the Motor by an ultrasonic cleaning apparatus.
The P200-WPD is made out of aluminum, therefore take special precaution to always use a PH neutral detergent
compatible with aluminum, when washing it in a Thermo Disinfector.
8-2-4 Sterilization
We verify the validity of sterilization in the following conditions.
1) Place it in a Primado Sterilization Cassette
2) Autoclave under the following conditions;
The sterilization parameters are indicated for sterilization and immediate use only.
CAUTION
Do not soak the Motor into water, or failure might occur.
The Motor is not suitable for EOG sterilization.
Sterilize the Motor after each patient.
Do not clean Motor by a washer sterilizer.
The Motors do not require lubrication. Never lubricate them.
8-3 Storage
Store the sterilized Motor in a place where it can be kept sterile until the next use.
9

Error Codes

When the Motor has stopped operating due to failure, overloading, breaking of a wire, misuse etc., the Control Unit displays Error Codes.
Then, an error code is displayed in the Mini Panel. If an error code is displayed, follow the check and measure steps shown in the display to
resolve the error. Touch the Close Button in the Sub Panel or operate the Motor or the Foot Control after resolving the error to close the error
display.
In case the error is unresolved, contact your dealer for repair.
If two or more error codes are involved, the error code displayed in the Mini Panel will blink. Touch the Mini Panel to