3-3 Connection of the Motor
By matching the mark
or B) at the front of the Control Unit, plug the Motor into the Connector all the way.
There are two Motor Connectors. By connecting two Motors to the two connectors, you can
select the one you want to use by the Motor Select Button (however, both Motors cannot be
operated simultaneously). The selected Motor can be confirmed in the Main Panel.
To disconnect the Motor, hold the Motor Cord Plug and pull it out with a straight movement.
Whenever you use two Motor Connectors alternately, be sure to confirm which Motor is under operation.
CAUTION
Do not connect anything else other than the Motor exclusively used for this system. Otherwise, the Motor Connector
might become deformed or failure could occur.
If the Motor Cord cannot be connected to the Control Unit easily, do not insist. Deformation or cracks might occur. A
foreign substance might be obstructing the connection or the connection portion might be deformed. Contact your
dealer.
Whenever connecting/disconnecting the Motor with Hand Control and Wire Pin Driver, be sure to switch OFF the Safety
Lock.
When connecting/disconnecting the Motor, do not turn the Motor Cord Plug. Otherwise, the inside pins might become
damaged, leading to failure.
Prevent the Motor Cord Plug from falling onto the floor or receiving a strong impact; otherwise, deformation might occur
and it might become impossible to establish a proper connection.
Do not connect/disconnect the Motor Cord with wet hands; otherwise, electric shock might occur.
3-4 Mounting of the Attachment/Handpiece
Whenever mounting/removing the Attachment/Handpiece, be sure to switch OFF the Safety Lock of the Wire Pin Driver/
CAUTION
Motor with Hand Control.
Prior to mounting/removing the Attachment/Handpiece, remove the Bur or Blade. Otherwise, injury might occur.
After mounting the Attachment/Handpiece, confirm that the Attachment/Handpiece and Motor are connected properly.
P200-HMH and P200-HMH-HS Attachments/Handpieces
<Mounting>
1) Push the Attachment/Handpiece into the Motor straight.
2) Turn the Attachment/Handpiece rightward or leftward so that the Alignment Pin of
the Motor matches the hole of the Attachment/Handpiece with a clicking sound to
connect firmly.
<Removing>
Pull out the Attachment/Handpiece from the Motor straight.
P200-SMH and P200-SMH-HS Attachments/Handpieces
<Mounting>
1) Push the Attachment/Handpiece into the Motor straight.
2) Align the '
then push the Attachment/Handpiece in until it clicks to connect firmly.
<Removing>
Pull out the Attachment/Handpiece from the Motor straight.
12
on the Motor Cord Plug and the mark
' Symbol of the Attachment/Handpiece and '
on the Motor Connector (A
' Symbol of the Motor,
Mark
Mark
Motor Cord Plug
Alignment Hole
Alignment Pin
Attachment /
Motor
Handpiece
Symbol
Symbol
Attachment /
Handpiece
Fig. 4
Fig. 5
Motor
Fig. 6