NSK Varios 560 LUX Operation Manual

NSK Varios 560 LUX Operation Manual

Multi function ultrasonic scaler

Advertisement

Multi Function Ultrasonic Scaler
OPERATION MANUAL
Please read this Operation Manual carefully
before use, and file for future reference.
The EU directive 93/42/EEC was applied in the design and production of this medical device.
OM-E0248E 004
'10.03.01 S

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Varios 560 LUX and is the answer not in the manual?

Questions and answers

Summary of Contents for NSK Varios 560 LUX

  • Page 1 Multi Function Ultrasonic Scaler OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference. The EU directive 93/42/EEC was applied in the design and production of this medical device. OM-E0248E 004 ’10.03.01 S...
  • Page 2: Cautions For Handling And Operation

    Original Operation Manual English Thank you for purchasing the Varios Ultrasonic Scaler (Varios 560 LUX / Varios 560). This product is used in dental office only. This device generates ultrasonic waves intended for use in dental applications such as scaling, root canal treatment, periodontal and cavity preparation. Read this operation manual carefully before use, and keep it within user's reach.
  • Page 3: Component Names

    · Do not force or pull on Power Cord and/or Handpiece Cord. It could cause disconnection. · Be sure to attach NSK genuine tips when using NSK Varios ultrasonic scaler (Varios 560 or Varios 560 LUX). The problems such as damage, failure and accident of handepieces resulting from use of Non-NSK tips are not included in 2.
  • Page 4: Installation And Assembly

    NOTICE Electrical Connections are made to the cord.) It might result in electrical shock. Fig. 7 * Picture shows Varios 560 LUX. Pushing the White Ring, (the Quick Connector Release Push Push Ring) on the Water Connector, gently remove the tube.
  • Page 5: Power Setting

    Fig. 13 CAUTION Model (6) Operate Varios 560 LUX, Varios 560 Check the case the selected Tip came in for proper power. · Depress the Foot Control. Tip will vibrate. (If the lamp stays illuminated by pushing the Water Key, water supply will You can check Maximum Power on the Tip Case Label.
  • Page 6 5. How to Use Tip Cover S (7) When Treatment is Finished. Handpiece Holder Handpiece Holder · Release the Foot Control and attach the handpiece to the When the Tip is mounted to the handpiece, install the Tip Cover handpiece holder. (Fig. 16) S for safety.
  • Page 7: Care And Maintenance

    2) Check wear of the Tip. one and reassemble the Filter in the reverse order. 3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we O-ring O-ring...
  • Page 8: Troubleshooting

    9. Troubleshooting (4) Changing Lamp (Varios 560 LUX) When trouble is found, check the following items again before asking repairs. If none of these are applicable or the trouble 1) Refer to 3. (5) Handpiece Connection. Disassemble the handpiece from the Handpiece Code. Remove the cover.
  • Page 9: Warranty

    Water is leaking from Maximum Output Firmly insert the Irrigation Tube into the the joint between the The Irrigation Tube is not Atmospheric pressure Varios 560 LUX : Max.35VA Irrigation Tube and the Irrigation Connector inmost. connected correctly. 700-1060 hPa Power Consumption Irrigation Connector.
  • Page 10 Guidance and manufacturer's declaration - electromagnetic immunity Cables and accessories Maximum length Shield Complies with The Varios560 / Varios560LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios560 / Varios560LUX RF emissions, CISPR11, Class B/ Group 1 Handpiece cord 2.0 m...