6-2 Adjusting the Motor
Touch the Motor Set Up Button in the Option Set Up screen to display the Motor Set Up screen.
Motor Set Up Screen
Acceleration Area
Oscillating Turns Area
6-2-1 Setting methods for Acceleration, Brake, Torque and Oscillating Turns
Push the UP/DOWN Button to set the desired value.
When two Motors are connected to Motor Connector A and B, the Motor to be set up can be selected by the Motor Select Button in the
Sub Panel.
The acceleration time for the Motor to reach its maximum rotational speed can be set at any level between 10% and 100%.
The time it takes for the Motor to stop can be set at any level between 10% and 100%.
The torque level of the Motor can be set between 30% and 100%.
The oscillating turns can be set between 1 and 5 (This is available only when the P200-WPD Wire Pin Driver is connected).
While the Motor Set Up screen is displayed, the Motor can be started. Prior to using the Motor, be sure to return to the
CAUTION
HOME screen.
6-2-2 How to return to the Factory Default settings
Touch the Factory Default Button and then the YES Button to
return to the Factory Default settings provided before delivery.
Touch the NO Button to keep the current settings.
6-3 Setting Up the Attachment
By setting up the connected Attachment/Handpiece, the optimal rotational speed for the particular Attachment/Handpiece becomes
available. After setting up the Attachment/Handpiece in the Attachment Set Up screen, return to the HOME screen to adjust the rotational
speed and the Irrigation volume.
When two Motor are connected to Motor Connector A and B, the Attachment/Handpiece to be set up can be selected by the Motor Select
Button in the Sub Panel.
Touch the Attachment Set Up Button in the Option Set Up screen to display the Attachment Set Up screen.
The settings of the Attachment/Handpiece are reset whenever the power is turned off or the Motor is disconnected from
CAUTION
the Control Unit. Prior to using the Attachment/Handpiece, be sure to confirm the current settings.
If the Motor is not connected, a message alert announcing that the connection is not made will be displayed in the Main
Panel.
Brake Area
Torque Area
Factory Default Button
Motor Select Button
Fig. 34
Fig. 35