Option CR3, or CRP, or ALV Installation Kit
Utilities, Adjustments, and Performance Verification Tests
Calibration Software and specified test equipment is required to perform the adjustments, and can
be used to automate the performance verification testing. Information on how to obtain this
software can be found at
http://www.keysight.com/find/calibrationsoftware
Utilities Required
None
Adjustments Required
If the instrument is more than one year old, or if this instrument has been adjusted as part of a
calibration or repair during its first year, the following adjustments are the minimum set required to
ensure that this newly installed option is functioning properly.
Front End Controller Attenuator
Performance Testing Required
None
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your