Performing the
Micro Fluidic
Card Run
26
June 23, 2003 9:39 am, MFC_QSG_text.fm
Additional Information About Creating SDS Templates and
Documents
The user guide for the 7900HT system provides information about:
• Creating SDS documents and templates manually
• Creating SDS documents and templates for absolute
quantification studies
Before performing the run, place the Micro Fluidic Card in the
7900HT system as explained in the following procedure.
The procedure assumes that you have imported the SDS setup
Note:
file and created an appropriate Micro Fluidic Card SDS document.
Refer to
"Creating Micro Fluidic Card Documents" on page 20
more information. Incorrect setup can lead to erroneous analysis or
loss of data.
To place the Micro Fluidic Card in the
7900HT system:
1. In the SDS 2.1 software, open the Micro Fluidic Card
document for the card you are running.
2. Click the Instrument tab.
3. In the Real-Time tab, click Open/Close.
The instrument tray rotates to the OUT position.
Applied Biosystems Micro Fluidic Card Getting Started Guide