Auxiliary fresh-gas outlet
Hide thumbs Also See for Fabius CE:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11
Using the external fresh gas outlet with an additional switch (optional)
The changeover switch permits simple switching of
the fresh gas from the compact breathing system
(COSY) to the non-rebreathing system.
Preparation
Example: Bain non-rebreathing system
2
2
1 Connect the non-rebreathing system to the
corresponding fresh gas outlet on the switch.
2 Screw the sample line to the connections on the
breathing mask or the breathing system filter
and to the water trap on the gas monitor.
3
3 If required, connect the anesthetic gas
scavenging hose of the non-rebreathing system
to the second port of the anesthetic gas
receiving system (AGS).
1
Supplement to Instructions for Use
Follow the Instructions for Use included with the
non-rebreathing system and the anesthetic gas
receiving system (AGS).
Operation with non-rebreathing system
Fresh gas flow to the non-rebreathing system:
Set the lever on the switch to »
The lever points in the direction of the non-
rebreathing system and shows the symbol for
the non-rebreathing system.
Set the fresh gas flow according to the patient's
requirements.
Operate the non-rebreathing system according
to the corresponding Instructions for Use.
«.