Supplement to Instructions for Use
General safety information
WARNING
Danger to the patient caused by too high
pressures
To ensure the correct inspiratory pressure at
the inspiratory port, use only non-rebreathing
systems with a reservoir bag conforming to
ISO 5362 and/or a pressure relief valve.
WARNING
Insufficient gas supply to the patient
Non-rebreathing systems are suitable and
intended only for manual ventilation or
spontaneous breathing and may only be
connected to the external fresh gas outlet.
WARNING
Risk of misinterpreting the measurements
The measurements and alarms are displayed
on the breathing monitor. They do not
correspond to the measurements for the
patient connected to the external fresh gas
outlet.
– The measurements (O
and alarms displayed on the breathing
monitor correspond to the measurements
on the compact breathing system (COSY).
– Switch to the Standby mode when using
the external fresh gas outlet.
WARNING
Danger to the patient
There is no monitoring of pressure, volume or
O
for the external fresh gas outlet.
2
An external monitoring unit must be used to
perform the required monitoring of O
anesthetic gas, volume and pressure.
12
, pressure, volume)
2
, CO
,
2
2
NOTE
The connected system (e.g. Magill, Kuhn, Waters,
Bain) must be operated in accordance with the
respective associated Instructions for Use.