Bacharach Fyrite InTech 0024-8510 Configuration And Operation Manual page 39

Combustion gas analyzer
Table of Contents

Advertisement

1.800.561.8187
Configuration
Setup Menu
34
www.
Fyrite
Function
value (in 100 ppm increments) before returning to the
Main Menu.
During combustion analysis, if the Protect CO feature is
enabled and the CO reading equals or exceeds the CO
Limit setting, the analyzer shuts off the pump and
prompts the operator for a course of action.
Continue (and risk sensor damage)
Purge
Continue Option:
Pump starts.
Future High CO warnings are suppressed until CO
drops below 100 ppm.
After CO drops below 100 ppm, the analyzer begins
enforcing the current Protect CO threshold limit
again.
Purge Option:
Remove probe tip from flue and press ENTER.
Pump starts.
Purge begins.
Measured CO is displayed.
Purge ends when CO level drops below 50 ppm.
The Combustion Air Temp option allows you to select a
source for the combustion air temperature (T-Air) used
by the analyzer for calculations. There are 3 choices.
1: T-Air value is read
dynamically from the
temperature sensor inside
the analyzer.
2: Temporarily use probe
(T-STK) to get T-AIR value.
When ready, press ENTER
to save displayed reading.
Value is saved and used as
fixed T-AIR value. After
save, the probe
(thermocouple) is used for
taking T-STK readings
again.
.com
InTech
Manual
0024-9486 Rev 4