Function
IMPORTANT:
Changing this setting
resets several configuration parameters
to their default values. Below is a list of
affected
parameters,
unaffected.
Reset to
Default Values
Temperature units
Manual/Auto zero
O
(Oxygen) reference
Calibration data
2
Fuel
User name
Memory erased
Clock
The CO Protect feature protects the CO sensor from the
negative effects of being overloaded with CO gas. Such
negative effects include:
a longer sensor recovery time
a shortened sensor life.
The CO sensor is an electro-chemical sensor whose
lifespan is a function of its exposure to the target gas
(CO in this case). Though the CO sensor is designed to
be used in combustion environments having a fairly
wide range of CO gas present, limiting unnecessary
overexposure to CO gas can greatly increase the life
expectancy of the CO sensor. In addition, it can shorten
sensor recovery time after exposure.
The Protect CO feature sets a protection threshold
above which the analyzer's pump shuts off, limiting the
CO sensor's exposure to the high levels of CO gas and its
negative effects. Enter the Protect CO screen from the
Setup Menu. Select the "ON" option and press the
ENTER key. The CO threshold limit is displayed. Use the
up and down arrows to modify the CO limit threshold
.com
Configuration
and
those
Unchanged