Combustion gas analyzer
Table of Contents

Advertisement

1.800.561.8187
Overview
Measure
Range
-ment
O
0 to 20.9 %
2
CO
0 to 2000 ppm
Stack
-20 to 650 C
Temp
(-4 to 1202 F)
Calculation
Calculation Range
Efficiency (HHV)
ETA (LHV)
Excess Air
Stack Loss
Lambda
CO
(dry basis)
0.1 to a fuel-dependent max in %
2
CO Ref to O
2
CO/CO
Ratio
0.0001 to fuel-dependent max
2
12
www.
Fyrite
Resolution
Accuracy
0.3% O
0.1% O2
(on flue gas)
2
10 ppm
(0 to 200)
1 ppm
5% reading (201 to 2000)
2 C (0 to 124 C)
1 C
3 C (125 to 249 C)
(1 F)
4 C (250 to 400 C)
Reso-
lution
0.1 to 100 %
0.1%
0 to 115%
0.1%
1 to 250 %
1%
0.1 to 100 %
0.1 %
1 to 9.55
0.01
0.1 %
0 to 9999 ppm
1 ppm
0.0001
  
.com
InTech
Manual
Response
Time
(T90)
< 20 sec
< 40 sec
< 50 sec
Version
NA
Siegert
X
X
X
X
X
X
X
X
X
X
X
0024-9486 Rev 4