2 Intended use
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
Approved for a body weight of up to 100 kg.
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
In general, all modular adapters are tested by the manufacturer for 3 million
load cycles. Depending on the patient's level of activity, this corresponds to
a service life of 3 to 5 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the prosthetic components according to the classification (see
section "Area of application").
Allowable single axis feet
1H* (requires 2S88 dorsal stop set), 1G9*
1H*