Download Print this page

Oscilloscope Setup - Keysight Technologies B2960A Demo Manual

Using an external trigger to generate pulses with 6.5 digit low noise power source

Advertisement

04 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide
Setup
Connection
1. Connect the black banana cable of the U8201A to the B2960A's Ch1 Low Force Terminal.
2. Connect the red banana cable of the U8201A to the B2960A's Ch1 High Force Terminal.
3. Clip the end of the 1.1 kΩ resistor with the black alligator clip of the U8201A.
4. Clip the other end of the 1.1 kΩ resistor with the red alligator clip of the U8201A.
5. Probe across the 1.1 kΩ resistor to monitor the output using the oscilloscope's Ch1.
6. Connect the Pin 9 of the B2960A's Digital I/O Port to the external trigger signal source and
the oscilloscope's Ch2 using the BNC cables, the N1294A-031, wiring, etc.
To oscilloscope's input Ch 1
B2960A
1.1 kΩ
U8201A
Resistor
BNC Cable
N1294A-031
To oscilloscope's input Ch 2
External Trigger Signal
BNC Cable
BNC Cable

Oscilloscope setup

1. Set the Ch1 vertical scale to 500 mV/div.
2. Set the Ch2 vertical scale to 5 V/div.
3. Set the horizontal scale to 500 μs/div.

Advertisement