Download Print this page

Keysight Technologies B2960A Demo Manual page 11

Using an external trigger to generate pulses with 6.5 digit low noise power source

Advertisement

11 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide
6. (Optional) Abort the channel to stop waiting for an external trigger signal
If you set the source trigger count to INFINITY, or the number of external trigger signals
was less than the specified source trigger count, you will need to abort the process.
1) If you aren't on the top of the Function menu, press
the top level.
On the top level of the Function menu
In the middle level of the Function menu
2) Press
,
stop waiting for an external trigger signal.
Conclusion
The Keysight B2960A 6.5 Digit Low Noise Power Source is a bench-top power supply with
revolutionary capabilities and functions not previously available. It can source either voltage or
current with 6.5 digits of resolution while also monitoring them, which is essential for a variety
of measurement applications.
In addition to an arbitrary waveform generation function that allows it to source current/voltage
waveforms in a variety of formats, the B2960A series also has excellent pulsed signal output
function and the triggering capability via its digital I/O signal port.
Configuring the B2960A series properly enables you to use an external trigger to generate puls-
es to evaluate and analyze a device-under-test (DUT). The addition of an oscilloscope to the
test setup allows you to monitor and analyze the waveforms.
B2900 Precision Instrument Family
The B2900 family contains products that perform both precision sourcing and precision
measurement. www.keysight.com/find/b2900a
No tree structure can be seen
Some tree structure can be seen
,
, and then press
repeatedly to return to
to make the channel 1
(1) Press Trigger
(2) Press Abort
(3) Press Trans.
(4) Press Ch 1

Advertisement

loading