Using an external trigger to generate pulses with 6.5 digit low noise power source

Advertisement

09 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide
14) Or press
to set it to INFINITY if you'd like to continue to source pulses. If you
set the trigger count to INFINITY be sure you abort the process at the end of the test.
4. Disable an internal source wait time
The B2960A series has internal wait time parameters to adjust the timing to source and
measure in order to have the measurements made under the exact source condition as
specified. However, it is required to disable it to source pulses at the exact timing as
specified by external trigger signals
1) If you aren't on the top of the Function menu, press
the top level.
On the top level of the Function menu
In the middle level of the Function menu
2) Press
,
dialogue.
3) Press
and select
No tree structure can be seen
Some tree structure can be seen
, and then press
to open the Wait Control
(1) Press Config
(2) Press Common
(3) Press Wait
to disable the Source wait time.
repeatedly to return to