7 Reprocessing methods according to ISO 17664
▶ In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry after
completion of the cycle, and then grease it immediately with servicing
agents from the KaVo care system.

7.4 Drying

Manual Drying
▶ Blow off the outside and inside with compressed air until water drops
are no longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the ther‐
modisinfector.
▶ Follow the instructions for use of the thermodisinfector.