7 Reprocessing methods according to ISO 17664
▶ For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector.
▶ Directly after automated cleaning/disinfection, treat the medical de‐
vice with the care products and systems provided by KaVo.
7.2.3 Manual cleaning - internal
Validated manual internal cleaning (removal of residual protein) can only
be done with KaVo CLEANspray and KaVo DRYspray.
▶ Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three
times for 2 seconds each time. Remove the medical device from the
spray attachment and let the cleaner work for one minute.
▶ Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
45 / 74