7 Reprocessing methods according to ISO 17664
▶ Immediately after internal disinfection, lubricate the KaVo medical de‐
vice immediately with care agents from the KaVo care system.

7.3.3 Machine disinfection - external and internal

KaVo recommends thermodisinfectors in accordance with EN ISO
15883-1, which are operated with alkaline cleaning agents with a pH value
of max. 10 (e.g. Miele G 7781 / G 7881 – Validation was carried out with
Programme "VARIO-TD", cleaning agent "neodisher® mediclean", neutral‐
isation agent "neodisher® Z" and rinsing agent "neodisher® mielclear"
and only applies to the material compatibility with KaVo products).
▶ For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector
(complying with max. pH value of 10).