Automatic Zero Setting At Power-On; Zero Tracking; Setting Units Of Measure; Semi-Automatic Tare (Net/Gross) - LAUMAS TLU Installation And User Manual

Table of Contents

Advertisement

 (from 0 to max 20% of full scale; default: 0): If at switch-on the weight value is lower than
the value set in this parameter and does not exceed the  value, the weight is reset. To
disable this function, set 0.
 (from 1 to 5, default: ): When the weight value is stable and, after a second, it
deviates from zero by a figure in divisions smaller or equal to the figure in divisions set in this
parameter, the weight is set to zero. To disable this function, set .
Example: if the parameter  is set to 5 and   is set to 2, the weight will be automatically
set to zero for variations smaller than or equal to 10 ( x  ).
These are the available units of measure:
kilograms
:
:
grams
tons
:
THE SEMI-AUTOMATIC TARE OPERATION IS LOST UPON INSTRUMENT POWER-
OFF.
To perform a net operation (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press the
TARE f key for less than 3 seconds. The instrument displays the net weight (just set to zero) and the
NET LED lights up. To display the gross weight again, keep the NET/GROSS input closed or press
TARE for 3 seconds.
This operation can be repeated many times by the operator to allow the loading of several products.
Example:
Put the box on the scale, the display shows the box weight; press TARE, the display shows the net
weight to zero; introduce the product in the box, the display shows the product weight. This
operation can be repeated several times.

AUTOMATIC ZERO SETTING AT POWER-ON

ZERO TRACKING

SETTING UNITS OF MEASURE

SEMI-AUTOMATIC TARE (NET/GROSS)

- 14 -