- It is recommended that the power supply negative pole be grounded.
- It is possible to supply up to eight 350 ohm load cells or sixteen 700 ohm load cells.
- For 4-wire load cells, make a jumper between EX- and REF- and between EX+ and REF+.
- Connect terminal 0 VDC" to the RS485 common of the connected instruments in the event that
these receive alternating current input or that they have an optically isolated RS485.
- In case of an RS485 network with several devices it is recommended to activate the 120 ohm
termination resistance on the two devices located at the ends of the network, as described in the
paragraph RS485 SERIAL CONNECTION.
- The analog board (ANALOG) is present on TLUANA model.
RS485 termination
J2
J1
OPTION ANALOG
IN 1: zero input
IN 2: NET/GROSS input

ELECTRICAL CONNECTIONS

BASIC INFORMATION

WIRING DIAGRAM

RS485
OPTION
ANALOG
RS232
OUTPUT
RX
22
23
24
25
26
27
28
29
30
OUT1
OUT2 OUT3 OUT4
1 2 3 4 5 6 7 8 9
RELAY OUTPUTS
115Vac 150mA
DIG. IN 12/24Volt
12-24Vdc
TX
31
32
33
NET
0
0
ESC
POWER
TLU
LOAD CELLS
10
1112
13
141516 1718 19 20
LOAD CELLS
- 6 -
kg
g
L
4-WIRE load cell
CONNECTION
16
17
21
6-WIRE load cell
CONNECTION
18
19
20
21