Volt:protection{<Vaule>|Minimum|Maximum; Volt:protection:[State] {0|1|Off|On - ITech IT6800B Series Programming Manual

Dc programmable power supply
Hide thumbs Also See for IT6800B Series:
Table of Contents

Advertisement

Parameters
MIN to MAX
*RST value
the stepping resolution of voltage of this unit.
Query order
VOLT:STEP?
Return parameters
<NR2>
VOLT:PROTection{<vaule>|MINimum|MAXimum}
This order is used to set the upper limit of over voltage protection point. If the
output voltage value is higher than the upper limit of OVP, then we can assume
that internal parts has been shorted. Query state register "OV" bit has been set.
When power supply is protected, then we can through order VOLT:PROT:CLE
to clear the OVP state.
Order syntax
[SOURce:]VOLTage:PROTection[:LEVel] <NRf>
Parameters
MIN to MAX
Unit
V
Query syntax
VOLT:PROT? { MINimum|MAXimum}
Return parameters
<NR2>

VOLT:PROTection:[STATe] {0|1|OFF|ON}

This order is used to open or close OVP function i.e to set the state of OVP.
Order syntax
[SOURce:]VOLTage:PROTection:STATe
Parameters
0|1|OFF|ON
Query syntax
VOLTage:PROTection:[STATe]?