Gross/Net Toggling - Sartorius EC1XS Operating Instructions Manual

Electronic stainless steel scales for use in hazardous areas
Table of Contents

Advertisement

Example
Toggle weight units from kilograms [kg] (1st weight unit) to Troy ounces [ozt] (2nd weight unit)
Settings (changes in the factory settings required for this example):
Application program: Toggling between weight units (2 1 2)
Weight unit 2: pounds (3 1 7)
Step
1. Switch on scale
2. Place load on scale
(in this example: 22.95 kg)
3. View weight in pounds [lb]
4. Print result
5. View weight in kilograms [kg]

Gross/Net Toggling

Purpose
Use this program to toggle the scale
between net and gross values.
Features
- Toggle between display of net value
and of gross value (when tare value is
in memory)
- Menu code 2 1 9
Preparation
Select "Gross/net toggling" in the oper-
ating menu as follows:
$ To activate the operating menu, switch
off the scale by pressing e [On/Off/
Standby]; switch the scale back on (e
[On/Off/Standby]) and while all seg-
ments are lit, press the ) [Tare] key
briefly.
§ Select the "Gross/net toggling" menu
item:
Press ) [Tare], p [Print], p
[Print], ) [Tare], repeatedly as needed
> Gross/net toggling menu code: 2 1 9
Key (or instruction)
e [On/Off/Standby] key
k [Function] key
p [Print] key
k [Function] key
§ Confirm selection: press and hold p
[Print] key (2 sec)
> "o" indicates the active setting
§ Set other parameters: Press the p
[Print] key
§ Save settings and close the menu: press
and hold the ) [Tare] key (2 sec)
Example of a printout
Net value (p [Print] key):
N
+
1250.0 kg
T
+
150.0 kg
G#
+
1400.0 kg
Gross value (p [Print] key):
G
+
1400.0 kg
Display/Data output
+
229.5 kg
+
506.0 lb
ACE HARDWARE
GOETTINGEN
G + 506.0 lb
+
229.5 kg