3x3 MIMO Configuration in Two PXIe Chassis with External Controller
For cabling details of the M9381A and M9391A refer to
For the first (top) configuration - slave
From the M9300A (slot 10) use three 8121-2175 cables to make the following connections:
100 MHz Out 1 to M9301A 100 MHz In (slot 9)
l
100 MHz Out 2 to M9301A 100 MHz In (slot 14)
l
100 MHz Out 3 to M9301A 100 MHz In (slot 5)
l
For the second (bottom) configuration - master
From the M9300A (slot 10) use three 8121-2175 cables to make the following connections:
100 MHz Out 1 to M9301A 100 MHz In (slot 11)
l
100 MHz Out 2 to M9301A 100 MHz In (slot 7)
l
100 MHz Out 4 to M9301A 100 MHz In (slot 4)
l
These configurations can be used for x2 or x3 source only or analyzer only configurations.
LTE/LTE-A Multi-Channel Reference Solution
MIMO Instrument Connections (page
Step 3: Install the Modules
17).
23