Features and Functions

External Trigger Out

Trigger Out selects one of the eight lines from trigger bus (TRIG [0..7]) as an
external trigger source. The selection of the trigger out line is done by the USB
device in the chassis as illustrated in
Table 2-2
and U2600A Series DAQ, whereas for U2700A Series modular products, user is
allowed to choose any trigger line from the trigger bus (TRIG [0..7]) as an external
trigger source.
The SCPI command below is used to select one of the lines or bits of the trigger
bus (TRIG [0..7]) as an external trigger source:
Table 2-2
Trigger out bits for U2300A, U2500A, and U2600A Series DAQ devices
Trigger Out
Bit-0
Bit-1
Bit-2
Bit-3
Bit-4
Bit-5
Bit-6
Bit-7
28
defines the available trigger out signals provided by U2300A, U2500A,
TRIGger:OUT {0|1|2|3|4|5|6|7}
Figure 2-2
by means of a multiplexer.
Function
Time base
Reserved
Reserved
A/D trigger
Reserved
Reserved
Reserved
D/A trigger