Troubleshooting Table

Symptom
Possible Cause
Oxygen Tubing
1
Kinked/Leaking
Humidifier Bottle
2
Restriction (if used)
3 Flow Meter Closed (off)
No Oxygen Flow
Internal Tubing
4
Kinked/Leaking
5 HEPA Filter Restriction
6 Compressor Malfunction
Oxygen Tubing
1
Kinked/Leaking
Humidifier Bottle
2
Restriction (if used)
Low or Fluctuat-
3 Poor Concentrator Location
ing Oxygen
Flow Rates
4 Intake Filter Restriction
5 HEPA Filter Restriction
Internal Tubing
6
Kinked/Leaking
7 Flow Meter Malfunction
8 Regulator Malfunction
9 Compressor Malfunction
1 Intake Filter Restriction
2 HEPA Filter Restriction
3 Poor Concentrator Location
Low Oxygen
Internal Tubing
4
Concentration
Kinked/Leaking
5 Solenoid Valve Malfunction
6 Compressor Malfunction
7 Sieve Bed Failure
1 Poor Concentrator Location
Overheating
2 Cooling Fan Malfunction
Internal Tubing
Outlet Pressure
1
Kinked/Leaking
Out of
Acceptable
2 Regulator Out of Adjustment
Range (4.3 - 4.9
3 Regulator Malfunction
PSI)
4 Sieve Bed Failure
Oxygen Concentrator
Check the nasal cannula and any extension tubing being used for kinks, blockages, or
leaks. Verify that all tubing connections are secure. Replace and tubing if necessary.
Check the humidifier bottle and tubing for blockages or restrictions.
Inspect the flow meter and verify that the ball is not at the bottom of the meter. If it is,
turn the knob to increase the flow. Replace the flow meter if necessary.
Check all internal tubing for leaks. Verify that there are no leaks and that all
connections are secure. Replace any tubing or parts if necessary.
Inspect the HEPA filter for a blockage. Replace if found to be clogged.
Verify that the compressor is operating properly. Replace the compressor if necessary.
Check the nasal cannula and any extension tubing being used for kinks, blockages, or
leaks. Verify that all tubing connections are secure. Replace any tubing if necessary.
Check the humidifier bottle and tubing for blockages or restrictions.
Verify that the concentrator is in a well-ventilated location and that air flow into
the device is not impeded.
Inspect the intake filter for a blockage. Replace if found to be clogged.
Inspect the HEPA filter for a blockage. Replace if found to be clogged.
Check all internal tubing for leaks. Verify that there are no leaks and that all connections
are secure. Replace any tubing or parts if necessary.
Verify the functionality and accuracy of the flow meter. Replace if necessary.
Check the outlet pressure. If it is out of the acceptable range, adjust or replace
the regulator.
Verify that the compressor is operating properly. Replace if necessary.
Inspect the intake filter for a blockage. Replace if found to be clogged.
Inspect the HEPA filter for a blockage. Replace if found to be clogged.
Verify that the concentrator is in a well-ventilated location and that air flow into
the device is not impeded.
Check all internal tubing for leaks. Verify that there are no leaks and that
all connections are secure. Replace any tubing or parts if necessary.
Verify the operation of the 2-way solenoid valve. Replace if necessary.
Verify that the compressor is operating properly. Replace if necessary.
Replace the sieve beds assembly.
Verify that the concentrator is in a well-ventilated location and that air flow into
the device is not impeded.
Verify that the cooling fan is connected properly to the circuit board and check the
continuity of the connection. Replace the cooling fan if it is not functioning.
Check all internal tubing for leaks. Verify that there are no leaks and that all connections
are secure. Replace any tubing or parts if necessary.
Adjust the regulator.
Replace the regulator.
Replace the sieve beds assembly.
Corrective Action
24