11.2 Safety distances

Recommended safe distances between portable and mobile
high-frequency telecommunication devices (e.g. mobile phones) and the
Rated output of
HF device
in W
0.01
0.1
1
10
100
Contact Weinmann for further technical specifications. More details can also be
found in the hospital manual and in the service and repair guide.
The right to make design modifications is reserved.
VENTImotion
Safe distance as a function of transmission frequency
150 KHz - 80 MHz
0.12
0.37
1.17
3.69
11.67
in m
80 MHz – 800 MHz
0.12
0.37
1.17
3.69
11.67
Technical specifications
800 MHz – 2.5 GHz
0.24
0.74
2.34
7.38
23.33