10.2 Safety distances

Recommended safety distances between portable and mobile
HF telecommunications equipment (e.g. cell phones) and the BiLevel ST 22
Nominal output of
HF equipment
in W
0.01
0.1
1
10
100
Further technical data can be obtained on request from Weinmann and are listed in the
hospital manual and the Service and Repair Instructions.
Subject to design modifications
46
EN
Technical data
Safety distance dependent on transmission frequency
150 kHz - 80 MHz
0.04
0.11
0.35
1.10
3.50
in m
80 MHz - 800 MHz
0.04
0.11
0.35
1.10
3.50
800 MHz – 2.5 GHz
0.08
0.22
0.70
2.20
7.00