Pneumatic Diagram - Weinmann VENTImotion Series Description Of Unit And Operating Instructions For Patients

Table of Contents

Advertisement

Available languages

Available languages

Fine filter service life
Permitted humidity for operation
and storage
System resistance at an air flow of
60 l/min at the patient connection
opening

11.1 Pneumatic diagram

O pressure
Flow adjuster
2
source (0.5-6 bar)
up to max. 4 l/min
Ambient
Filter element
Fan
air inlet
Pressure sensor for
patient pressure
48
EN
Technical specifications
VENTImotion with hose system
WM 24130 and Silentflow
0.19
Vent to
ambient air
Optional
O valve
2
Flow sensor
Optional
Metering pipe
humidifier
Pressure sensor for
appliance pressure
VENTImotion
1000 hours with normal room air
≤ 95 % rH (no condensate formed)
WM 23600
kPa · s
l
Optional
Ventilation hose
bacteria filter
VENTImotion with VENTIclick
VENTImotion with O
hose system
2
WM 23737, VENTIclick WM 24365
and bacteria filter WM 24148
kPa · s
0.29
l
Expiration
1,8 m
system
Patient mask
(nasal or full
facial mask)