IMPORTANT SAFETY INSTRUCTIONS
IMPORTANT SAFETY INSTRUCTIONS
CAUTION
1.
The use of accessories, transducers, and cables other than those specified by the manufacturer, may result in increased emissions or
decreased immunity of the table.
2.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the table
should be observed to verify normal operation in the configuration in which it will be used.
3.
The table is designed to be a stand-alone table. This table must not be modified or incorporated into any other
equipment.
4.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in
the area of this equipment. Directions for use of this equipment are described in this manual. The operator should read these
sections carefully.
5.
Weight Limit: (patient and accessories) 500 lbs. (227 kg.) Do not exceed.
6.
Adjustable Body Comfort (ABC) System Weight Limit - 100 lbs. (45 kg.)
7.
Do not sit on the ABC System.
8.
Do not place foreign objects into the ABC System.
9.
Only use the ABC system as intended.
10. For safety of your client, use a cloth sheet between the ABC System and your client.
11. Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory
patient. The patient may lose balance and fall.
12. Do not lift table by the tabletop pad. This can damage the removable pad.
13. When lowering the table, make sure there is nothing underneath the tabletop that can impede motion (like stools,
cabinets, accessory parts, cleaners, etc...)
14. When using the table Height Up function, do not place anything between the metal frame and the base structure.
Damage can occur to the table.
15. Do not sit on backrest or footrest sections.
16. The hood area under the backrest top is not designed for storage.
17. DO NOT use the hood area under the backrest top for holding anything as damage can occur to the table.
18. Before moving any of the tables, disconnect the power cord from the power outlet.
19. The table weighs as much as 475 lbs. (216 kg.)
20. Use extreme caution when lifting the table to move it.
21. 4 people are required to move the table safely.
22. DO NOT use the table to transport a patient.