FRED PA-1
Step 3
Fig. 4.8
Button to deliver the shock
Step 4
Finishing the therapy
User Guide
Shock delivery
When the energy is charged, the user is prompted to trigger the shock by pressing the
blinking
orange button.
Shock hazard!
 Do not, under any circumstances, touch the patient during shock delivery.
 Make sure that the patient does not touch any conducting objects.
5. Deliver the shock by pressing the button
After the shock delivery, proceed with
citation.
Performing cardiopulmonary resuscitation
If the FreeCPR option is activated, the device instructs the rescuer to adjust the chest
compression frequency.
FreeCPR measures the compression rate based on the impedance measurement by
the defibrillation electrodes.
6.
Perform a CPR cycle. According to the configuration of the device, a CPR cycle
consists of:
– performing chest compressions for the set period of time, or
– alternately performing 30 chest compressions and 2 breathes for the set period
of time.
After the CPR cycle, the device continues automatically with
ECG
signal.
See
4.6 Finishing the
therapy.
Defibrillation
Semi-automatic defibrillation
.
Step 4 Performing cardiopulmonary resus-
Step 2 Analysing the
4
4.3
Page 33