User calibration
You can perform calibration by using your own reference plate and calibration data instead of white calibration.
The calibration data for user calibration can be specified by connecting the instrument to a computer and using
the optional Color Management Software "SpectraMagic NX". Users can select whether to use the user calibration
data for measurement.
Operating Procedure
1
Press [MENU].
The <User calibration> screen is displayed.
Settings
{ OFF : User calibration cannot be performed.
{ ON : User calibration can be performed.
2
Use [◀], [▲], [▶], or [▼] to move the cursor to
select "OFF" or "ON".
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Calibration setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Calibration setup> screen.
Start the procedure from the <Setting> - <Calibration setup> screen.
User calibration
OFF
ON
4