Konica Minolta CM-26dG Instruction Manual page 41

Hide thumbs Also See for CM-26dG:
Table of Contents

Advertisement

3
Set the instrument in the calibration stage in order to
measure the calibration stage's white calibration plate.
Calibration stage
Notes
• Verify that the White ID shown on the screen and the
calibration stage number match.
Calibration stage
4
Press the measurement button.
White calibration will be performed.
Notes
• Do not move the instrument until white calibration is
complete.
Calibration stage
After white calibration is complete, a screen prompting
gloss calibration will be displayed.
Calibration stage
5
Set the instrument in the calibration staged in order to
measure the calibration stage's gloss calibration plate.
Notes
• Gloss calibration is only available with the 26dG.
• Verify that the Gloss ID shown on the screen and the
Calibration stage
calibration stage number match.
Calibration stage
6
Press the measurement button.
Gloss calibration will be performed.
Notes
• Do not move the instrument until gloss calibration is
complete.
When gloss calibration is completed, the screen returns to
the <Sample> screen.
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Please set the
instrument on the white
calibration plate and
press the measurement
button.
White ID: xxxxxxx
CALIBRATION
Calibrating to white
Please set the
instrument on the gloss
calibration plate and
press the measurement
button.
Gloss ID: xxxxxxx
CALIBRATION
Gloss calibrating
2
38

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cm-26dCm-25d

Table of Contents