Measurement range 0 to 200 GU;
Measurement area MAV : 10×7 mm,
Repeatability
Inter-instrument
agreement
Applicable
standards
Measurement time
Minimum measurement
interval
Data memory
Battery performance
Viewfinder function
Display
Display language
Interface
Power
Charging time
Operating temperature/
humidity range
Storage Temperature/
humidity range
Size
Weight
*1 Firmware version 1.10 or later and optional Color Management Software SpectraMagic NX (Ver. 3.0 or later) is required to
use UV Adjusted setting.
*2 Optional Color Management Software SpectraMagic NX (Ver. 2.9 or later) is required for setting user-configured illuminants
or user indexes.
141
CM-26dG
Display resolution: 0.01 GU
SAV : Ø3 mm
Standard deviation
0 to 9.99 GU: Within 0.1 GU
10 to 99.99 GU: Within 0.2 GU
100 to 200 GU: Within 0.2%
of indicated value
(When measured 30 times
at 5-second intervals after
calibration)
0 to 9.99 GU: Within ±0.2 GU
10 to 99.99 GU: Within ±0.5 GU
(MAV; compared to values
measured with a master body
under KONICA MINOLTA
standard measurement
conditions)
JIS Z8741 (MAV only),
JIS K5600, ISO 2813, ISO 7668
(MAV only), ASTM D523-08,
ASTM D2457-13, DIN 67530
Approx. 1 sec.
(Measurement mode: SCI +
Gloss or SCE + Gloss)
(From pressing measuring button to measurement completion)
Approx. 2 sec
(Measurement mode: SCI +
gloss or SCE + gloss)
1,000 target data + 5,100 sample data
Measurement mode: SCI +
Gloss or SCE + Gloss
Approx. 3,000 measurements (approx. 1,000 measurements when using Bluetooth)
when measurements are taken at 10-second intervals at 23°C with the dedicated
lithium battery
Available (with white LED illumination)
2.7-inch TFT color LCD
English, Japanese , German, French, Italian, Spanish, Simplified Chinese, Portuguese,
Russian, Turkish, Polish
USB 2.0; Bluetooth (SPP-compatible. Optional Bluetooth module required)
Dedicated lithium-ion battery (removable), USB bus power (with lithium-ion battery
installed), Dedicated AC adapter (with lithium-ion battery installed)
Approx. 6 h
Temperature: 5 to 40°C, Relative humidity: 80% or less (at 35°C) with no condensation
Temperature: 0 to 45°C, Relative humidity: 80% or less (at 35°C) with no condensation
Approx. 81 (W) × 93 (H) × 229 (D) mm
Approx. 660 g
CM-26d
Approx. 0.7 sec. (Measurement mode: SCI or SCE)
Approx. 1.5 sec
(Measurement mode: SCI or SCE)
Measurement mode: SCI or SCE
Approx. 630 g
CM-25d
Approx. 620 g

Advertisement