10. Overheating might occur if the handpiece or accessory bearings are
worn or are not kept clean. Continually check all parts of the handpiece
and attachments for overheating. Discontinue use and return the
equipment for service as necessary. Overheating can cause serious
injury to the patient or operating room personnel.
11. Please pay close attention to the "Care and Cleaning Instructions" and
all "Cautions" in the " MicroChoice ™ System Instruction Manual ".
12. Never immerse the MicroChoice Controller, Footswitch or handpieces
in any solution.
13. Do not lubricate any MicroChoice handpiece.
®
14. Hall
Surgical equipment is designed for use by medical professionals
who are completely familiar with the required surgical techniques and
instructions for use of the equipment.
15. Do not autoclave the MicroChoice Controller.
16. Always inspect cords for signs of excessive wear or damage.
Discontinue use and replace immediately or return for repair.
17. Never lock the collet on the High Speed or Medium Speed Drills
without a bur in place, or damage to the collet may result.
18. Do not pick up MicroChoice handpieces or the Footswitch by the cord.
Damage to the connection or cord may result.
19. Do not press the Mode button on the Handpiece Cord until a handpiece
is attached.