Cochlear Nucleus Freedom Bodyworn User Manual page 34

Speech processor
Hide thumbs Also See for Nucleus Freedom Bodyworn:
Table of Contents

Advertisement

The sound signal from an amplifier, TV or Hi-Fi can also be sent from
an induction loop to the telecoil.
To select the in-built telecoil, push both the Increase and Decrease
buttons at the same time, for a few seconds, until a T shows on the
display.
Make sure your finger presses both sides of the button area.
To disable the telecoil when it is selected, push and hold the Increase
and Decrease buttons together for a few seconds.
Your clinician can set your program to operate as either a telecoil
alone, or as a mix of your in-built telecoil and processor microphone.
If your clinician sets your program as a mix of the telecoil and
microphone, this will show as MT on the display.
Sensitivity and/or volume levels can be changed when using T or
MT in the same way as described previously. This will increase or
decrease the sensitivity of the telecoil (as well as the microphone
if MT is used) and overall volume of the telecoil (as well as the
microphone if MT is used).
To reset your program after using T or MT push all three buttons
together (Increase, Decrease and Select) for a few seconds. Your
Processor will reset the program to microphone (M) as well as the
sensitivity and volume levels if they were adjusted.
Disable the telecoil when not using the telephone, as it can
reduce the microphone volume.
34 - Nucleus Freedom User Manual
Telecoil selected
Mix of telecoil and microphone
© Cochlear Limited 2014