Calibration Offset Function
Set calibration offset
3
Calibrate
(Ex.)
Indicated temperature on controller: 37℃
Measured temperature: 36℃
Calibration offset: -1
+2.5℃
TROUBLE
ENTER
CANCEL
↓
① Press the ENTER key. The setting is decided
and the function menu selection screen is
displayed.
Offset is adjustable within the range from +3.0℃ to
-3.0℃.
Temperature correction in
overheating prevention device:
The temperature of overheating prevention device is
automatically corrected by the same amount when
that of controller is collected.
43
Operation Method