Fixed Temperature Operation
Set temperature
3
4
Start operation
POWER
① The
setting
temperature
displayed.
The Sub Display indicates "Set
TEMP" and the numeric character that indicates
temperature blinks.
② Set the temperature using the "▲▼".
③ Press the ENTER key to decide the temperature
and start the fixed temperature operation.
CAUTION!
 The refrigerator operation can not be performed
again for five minutes just after an operation stop.
The refrigerator has been in the standby state and
the REFRIGE lamp blinks during the period.
① The blinking FIXED TEMP lamp lights on when
the fixed temperature operation starts. The unit
starts to control temperature according to the
setting temperature.
on when the heater is on and The REFRIGE lamp
lights on when the refrigerator is on.
② The Sub Display displays the setting temperature.
The
arrow
which
temperature control is also displayed with
blinking.
The direction of arrow shows as
follows depending on the relation between the
setting temperature at operation start and that in
furnace.
Set TEMP
37.0℃↑
2002/01/01
12:00
(When setting temperature is
higher than temperature in
furnace)
③ The direction of arrow shows as shown below
when the temperature in furnace reaches to
within ±1.0℃ of setting temperature.
Set TEMP
2002/01/01
(When temperature in furnace
reaches
temperature)
④ Press the POWER key to stop operation.
15
Operation Method
input
screen
The HEATER lamp lights
indicates
the
state
Set TEMP
37.0℃↓
2002/01/01
12:00
(When setting temperature is
lower
than
temperature
furnace)
37.0℃→
12:00
to
around
setting
is
of
in

Advertisement