Auto Stop Operation
4
Start operation
① Press the ENTER key to decide the setting and
the auto stop operation starts.
AUTO STOP lamp lights on and the Sub Display
displays the setting temperature and residual time
to operation stop.
② The countdown of timer is suspended when the
temperature in furnace is out of the range of
"within ± 1 ℃ to the setting temperature".
this case, the Sub Display displays "Wait" with
blinking. The time display on the right side of
"Wait" shows the total waiting time in operation.
Set TEMP
Wait
(Waiting screen)
③ The operation stops when the residual time
counts zero. The END lamp blinks and the Sub
Display displays the operation finish time when
the operation stops.
 The wait function is not activated when the auto stop
operation is carried out with "Clock" mode.
operation stops at specified time.
④ Press the POWER key to quit operation.