Yamato IN604W Instruction Manual page 36

Programmable low temperature incubator
Table of Contents

Advertisement

Program Creation Example
25
Set the "No" using the "▼▲".
Press the ENTER key to decide the setting.
26
Press the PROGRAM key until the "Set TEMP" in the Step 3 is
27
displayed. The display on the lower column changes from "S02"
to "S03"or to "SO3 (un-used)".
Press the ENTER key to display the setting temperature edition
28
screen. The cursor goes out and the setting temperature blinks.
29
Set the temperature to -10℃ using the "▼▲".
Press the ENTER key to decide the temperature.
30
Press the PROGRAM key to display the "Set time" selection
31
screen.
Press the ENTER key to display the setting period edition screen.
32
The cursor goes out and the setting period blinks.
33
Set the period to 15 min using the "▼▲".
Press the ENTER key to decide the period.
34
Press the PROGRAM key to display the "Rep. start".
35
Press the ENTER key to display the repeat initiating step edition
36
screen. The cursor goes out and the setting for repeat initiating
step blinks.
37
Set "S02" using the "▼▲".
Press the ENTER key to decide the setting.
38
33
Operation Method
► No
Rep. start
S02
► No
Rep. start
S02
► 0.0℃
Set TEMP
S03
Set TEMP
0.0℃
S03
Set TEMP
-10.0℃
S03
► -10.0℃
Set TEMP
S03
► 0min
Set time
S03
Set time
0min
S03
Set time
15min
S03
► 15min
Set time
S03
► No
Rep. start
S03
Rep. start
No
S03
Rep. start
S02
S03
► S02
Rep. start
S03

Advertisement

Table of Contents
loading

Table of Contents