24
:POWer Commands

:POWer:PSRR:FREQuency:SINGle

Command Syntax
:POWer:PSRR:FREQuency:SINGle <value>[suffix]
<value> ::= {1 | 10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
[suffix] ::= {Hz | kHz| MHz}
The :POWer:PSRR:FREQuency:SINGle command sets the single frequency value.
The measurement is displayed on a log scale, so you can select from decade
values.
Query Syntax
:POWer:PSRR:FREQuency:SINGle?
The :POWer:PSRR:FREQuency:SINGle query returns the single frequency setting.
Return Format
<value><NL>
<value> ::= {1 | 10 | 100 | 1000 | 10000 | 100000 | 1000000 | 10000000
See Also
650
(see
page
1368)
| 2000000}
| 2000000}
":POWer:PSRR"
on page 644
":POWer:PSRR:APPLy"
":POWer:PSRR:DATA"
":POWer:PSRR:FREQuency:MODE"
":POWer:PSRR:PPDecade"
":POWer:PSRR:SOURce:INPut"
":POWer:PSRR:SOURce:OUTPut"
":POWer:PSRR:WGEN:LOAD"
":POWer:PSRR:WGEN:VOLTage"
":POWer:PSRR:WGEN:VOLTage:PROFile"