Connecting
4.5 Connecting the external DSL
The sensor cables are delivered with a pre-mounted F-connector at transmitter end.
1. Remove blind plug from external DSL enclosure.
2. Slide the cable gland back onto the cable and allow access for the F-connector tool.
3. Place F-connector tool around cable and slide up to engage F-connector nut.
4. Push cable through gland opening. Ensure that center lead is aligned with connector port in
external DSL.
5. Tighten F-connector using tool until you meet mechanical stop.
6. Remove F-connector tool.
7. Mount and tighten cable gland.
Repeat these steps for each sensor cable.
4.5.3

Channel 5 and 6 input configuration

Note
Connection of optional devices
The external DSL provides the option of connecting two additional devices to channels 5 and
6. Do not connect two devices to one channel.
Note
Analog input on channel 6
Hardware version 2 and below does not support analog input on channel 6.