Option HL1 Microwave Preselector Bypass (MPB) and Low Noise Path (LNP) Retrofit Kit
13.Refer to
Figure
10. Install rigid cable N9020-20210 (1) between attenuator (connected to
center switch port 1 and bottom switch port 1. Torque to 10 inch-lbs.
Figure 10
N9020-20210, N9020-20144, and N9020-20164 Rigid Cables Installation
14.Install rigid cable N9020-20144 (2) from top switch port 2 to center switch center port.
Torque to 10 inch-lbs.
15.Install rigid cable N9020-20164 (3) from bottom switch center port to Front End Assembly J9.
Torque to 10 inch-lbs.